millipore除 菌过滤器MPGL1GCA3 MPGL1GCA3


millipore除 菌过滤器MPGL1GCA3

millipore除 菌过滤器MPGL1GCA3 Sterile Millipak®-100 Filter Unit 0.22µm 9/16in. HB/HB millipore除 菌过滤器MPGL1GCA3 0.22 µm

millipore除 菌过滤器MPGL1GCA3

Description
Catalogue Number MPGL1GCA3
Trade Name
  • Millipak®
Description Sterile Millipak®-100 Filter Unit 0.22µm 9/16in. HB/HB
Product Information
Device Configuration Gamma Gold Capsule
Connections, Inlet/Outlet 14mm (9/16in.) Hose Barb
Maximum Differential Pressure, bar (psid) Forward: 4.1 bar (60 psid) @ 25°C; 1.7 bar (25 psid) @ 80°C; 0.35 bar (5 psid) @ 123°C; Reverse: 0.7 bar (10 psid) @ 25°C
Maximum Inlet Pressure, bar (psi) 5.2 bar (75 psi) @ 25°C
Good Manufacturing Practices These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
Biological Information
Bacterial Endotoxins Aqueous extraction contains ≤0.5EU/mL as determined by the Limulus Amebocyte Lysate (LAL) Test
Media Durapore®
Sterility Sterile
Sterilization 3 autoclave cycles of 90min @ 123°C; not in-line steam sterilizable
Bacterial Retention Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷CFU/cm² using ASTM®F838 methodology.
Wettability Hydrophilic

Physicochemical Information
Pore Size 0.22 µm
Inlet Connection Hose Barb
Outlet Connection Hose Barb
Bubble Point at 23°C ≥3450mbar (50psig) air with water
Flow Rate 11 L/min @ 2.8bar ΔP
Gravimetric Extractables The extractables level was equal to or less than 2.5mg per sterilized device after a 200mL flush and 24hours in water at controlled room temperature.
Non-Fiber Releasing This product was manufactured with a Durapore® membrane which meets the criteria for a “non-fiber releasing” filter as defined in 21CFR210.3(b)(6).
USP Bacterial Endotoxins A sample aqueous extraction contains less than 0.5EU/mL as determined by the Limulus Amebocyte Lysate (LAL) test.

Dimensions
Filtration Area 500 cm²
Process Volume 50 L
Inlet Connection Diameter 9/16 in.
Outlet Connection Diameter 9/16 in.

Materials Information
Chemistry
  • Polyvinylidene Fluoride (PVDF)
Device Material
  • Polycarbonate
Support Material Polycarbonate
Vent Cap Material Polyvinylidene fluoride (PVDF)

Packaging Information
Material Size 3
Material Package Blister

Specifications
Oxidizable Substances Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥200mL