millipore滤芯CVHL71TP3 CVHL71TP3


millipore滤芯CVHL71TP3

millipore滤芯CVHL71TP3 Durapore® Cartridge Filter with prefilter 10 in. 0.45 µm Code 7 millipore滤芯CVHL71TP3 双易打开袋 3个装

millipore滤芯CVHL71TP3

Durapore® Cartridge Filter with prefilter 10 in. 0.45 µm Code 7

Description
Catalogue Number CVHL71TP3
Trade Name
  • Durapore®
Description Durapore® Cartridge Filter with prefilter 10in. 0.45µm Code 7
Product Information
Device Configuration Cartridge
Maximum Differential Pressure, bar (psid) Forward: 5.5 bar (80 psid) @ 25°C; 1.7 bar (25 psid) @ 80°C; 0.35 bar (5 psid) @ 135°C. Reverse: 3.5 bar (50 psid) @ 25°C
Maximum Inlet Pressure, bar (psi) Forward- 5.5 bar (80 psi) @ 25°C; 1.7 bar (25 psi) @ 80°C; 0.35 bar (5 psi) @ 135°C; Reverse- 3.5 bar (50 psi) @ 25°C
Cartridge Code Code 7 (2-226) O-rings w/locking-tabs
Good Manufacturing Practices These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
Prefilter Material Mixed Cellulose Esters (MCE)

Biological Information
Bacterial Endotoxins Aqueous extraction contains ≤0.5EU/mL as determined by the Limulus Amebocyte Lysate (LAL) Test
Media Durapore® w/Prefilter
Sterility Other
Sterilization 10 SIP cycles of 30min or 10 autoclave cycles of 60min @ 121°C
Wettability Hydrophilic

Physicochemical Information
Pore Size 0.45 µm
Air Diffusion at 23°C ≤15mL/min @ 1.5bar (22psig) in water
Bubble Point at 23°C ≥1930mbar (28psig) air with water
Gravimetric Extractables The extractables level was equal to or less than 45mg per 10-inch cartridge after 24hours in water at controlled room temperature.
Indirect Food Additive All component materials meet the FDA Indirect Food Additive requirements cited in 21CFR177-182. All component materials meet the requirements of the EU framework regulation [1935/2004/EC] regarding materials and articles intended to contact food.
Non-Fiber Releasing This product was manufactured with a Durapore® membrane which meets the criteria for a “non-fiber releasing” filter as defined in 21CFR210.3(b)(6).
USP Bacterial Endotoxins A sample aqueous extraction contains less than 0.5EU/mL as determined by the Limulus Amebocyte Lysate (LAL) test.

Dimensions
Cartridge Nominal Length 10 in. (25 cm)
Diameter 6.9 cm (2.7 in.)
Filtration Area 0.69 m²
Device Size 10 in.

Materials Information
Chemistry
  • Polyvinylidene Fluoride (PVDF)
Device Material
  • Polypropylene
Seal Material Silicone (SI)
Support Material Polypropylene